White 23-liter class B autoclave with USB and integrated printer
Warranty Details
Safety
- It is essential that the water quality is below 20 microsiemens for the autoclave to start cycles.
- The cycle does not start if the door is not closed properly.
- The automated double hydraulic interlocking safety device prevents the door from opening while the chamber is under pressure.
- Safety valves protect the pressure generator and the chamber.
- A safety thermostat prevents overheating in the chamber and the generator.
- An automatic safety lock prevents accidental opening of the door during the sterilization process; the protector automatically cuts off the electrical power if there is a short circuit or if the power supply is unstable.
- Malfunction alarm indicates a failed cycle and a detailed error code.
- Door alarm indicates that the door is not secured.
- Low water alarm indicates when the water tank is empty or overfilled.
Technical Specifications
- Voltage: 220-240 VAC; 50/60 Hz; 10A
- External dimensions of the unit: W x H x L: 490 x 455 x 690 mm
- Tray dimensions: 380 x 192 mm
- Circuit breaker: F16A / 400V
- Rated power: 1500 VA
- 1750W
- Sterilization temperature: 121 ° C / 134 ° C
- Distilled water tank capacity: 2.5 L
- Operating temperature: 5 ° C / 40
- Relative humidity: Max. 80%, non-condensing
- Net weight: 53 kg
- Maximum noise level: <70 dB
- Atmospheric pressure of 76 kPa / 106 kPa
Icanclave Autoclave: maintenance and error message guide
Components
- Distilled water tank
- LCD screen
- Control panel
- Main switch
- Hose for draining distilled and used water
- Key for tightening the door
- USB port
- Internal printer (allows printing of sterilization cycle reports on thermal paper after the cycle ends.)
- Safety valve
- Ventilation grille
- Locking system
- Power cable
- Label
Included Accessories
- Sterilizer
- USB
- 4 instrument trays
- Tray support
- Handle for tray extraction
- Tool to adjust the door
- 2 drain hoses
- Door gasket
- Instruction manual
Regulations
- 93/42/EEC Medical Devices Directive.
- 97/23/EEC Pressure Equipment Directive.
- EN 13060 Small steam sterilizers.
- EN 61326, EN 61000 EMC.
- EN 61010-1, UL 61010-1 Safety requirements.
- EN 61010-2-040, IEC 61010-2-040 Particular requirements for steam sterilizers used for the treatment of medical materials.
Autoclave process step by step
1. Open the door, take out the tools and accessories located in the sterilization chamber, remove the packaging, and clean them.
2. Connect the power cord to a suitable power source.
3. Connect the printer.
4. Power on: the main switch is located on the front, bottom right of the unit. Once turned on, the LCD screen will light up, and the basic operating symbols of the sterilizer, such as door position, water level, sterilization program, date, and time will illuminate.
5. Remove the top cover then fill the distilled water tank. If a beeping sound occurs, it means you have exceeded the permitted water level in the tank. Drain the excess water until the alarm turns off.
6. Preparation of material to be sterilized: place the instruments on the different trays, and separate them according to their materials. If the instruments are not made of stainless steel, use sterilization pouches. These must not be in contact. Ensure all instruments are properly placed so that water does not remain in hollow parts. Do not overload instrument trays and do not place them in direct contact with the walls of the sterilization chamber. Do not use paper clips or needles to avoid any damage or malfunction of the autoclave.
7. Select the sterilization program.
8. Start the sterilization cycle.
After selecting the sterilization program, place the instruments on the trays in the chamber using the supplied plastic clamp.
Once the instruments are properly placed inside, close the door. Caution: Make sure the door is closed properly, otherwise the sterilizer will not operate.
9. Start the sterilization program.
Press the start button, the autoclave will begin operating for approximately 30-75 minutes.
10. End of the sterilization cycle.
Once the sterilization cycle is complete, the printer will automatically print the full cycle report.
Once the pressure has dropped to 0, you can open the door and remove the instruments. The autoclave will notify you with an audible signal once the sterilization cycle is complete.
Caution: Use the plastic clamp to remove the instruments.
IMPORTANT: Do not reuse wastewater after draining the tank.
It is important to keep the original packaging in case of after-sales service